Microfocus/Polycapillary-Optic Crystallographic X-Ray System

This system generates an intense, nearly collimated beam suitable for crystallography.

A system that generates an intense, nearly collimated, nearly monochromatic, small-diameter x-ray beam has been developed for use in macromolecular crystallography. A conventional x-ray system for macromolecular crystallography includes a rotatinganode x-ray source, which is massive (500 kg), large (approximately 2 by 2 by 1 m), and power-hungry (between 2 and 18 kW). In contrast, the present system generates a beam of the required brightness from a microfocus source, which is small and light enough to be mounted on a laboratory bench, and operates at a power level of only tens of watts.

ImageThe figure schematically depicts the system as configured for observing x-ray diffraction from a macromolecular crystal. In addition to the microfocus x-ray source, the system includes a polycapillary optic — a monolithic block (typically a bundle of fused glass tubes) that contains thousands of straight or gently curved capillary channels, along which x-rays propagate with multiple reflections. This particular polycapillary optic is configured to act as a collimator; the x-ray beam that emerges from its output face consists of quasi-parallel subbeams with a small angular divergence and a diameter comparable to the size of a crystal to be studied. The gap between the microfocus x-ray source and the input face of the polycapillary optic is chosen consistently with the focal length of the polycapillary optic and the need to maximize the solid angle subtended by the optic in order to maximize the collimated x-ray flux. The spectrum from the source contains a significant component of Cu K (photon energy is 8.08 keV) radiation. The beam is monochromatized (for Cu K) by a nickel filter 10 µm thick.

In a test, this system was operated at a power of 40 W (current of 897 µA at an accelerating potential of 45 kV), with an anode x-ray spot size of 41±2 µm. Also tested, in order to provide a standard for comparison, was a commercial rotatinganode x-ray crystallographic system with a pyrolytic graphite monochromator and a 250-µm pinhole collimator, operating at a power of 3.15 kW (current of 70 mA at an accelerating potential of 45 kV).

The flux of collimated Cu K radiation in this system was found to be approximately 16 times that in the rotatinganode system. Data on x-ray diffraction from crystals of tetragonal form of lysozyme (protein) in this system were found to be of high quality and to be reducible by use of standard crystallographic software.

This work was done by Marshall Joy of Marshall Space Flight Center, Mikhail Gubarev of the National Research Council, and Ewa Ciszak of Universities Space Research Association.

This invention is owned by NASA, and a patent application has been filed. For further information, contact Jim Dowdy, MSFC Commercialization Assistance Lead, at This email address is being protected from spambots. You need JavaScript enabled to view it.">This email address is being protected from spambots. You need JavaScript enabled to view it.. Refer to MFS-31499.

 

 