Coaxial electric heaters have been conceived for use in highly sensitive instruments in which there are requirements for compact heaters but stray magnetic fields associated with heater electric currents would adversely affect operation. Such instruments include atomic clocks and magnetometers that utilize heated atomic-sample cells, wherein stray magnetic fields at picotesla levels could introduce systematic errors into instrument readings.

Figure 1. A Typical Coaxial Electric Heater as shown in cross section resembles conventional signal-transmission coaxial cables, except that its outer conductor is deliberately made highly resistive.
A coaxial electric heater (see Figure 1) is essentially an axisymmetric coaxial cable, the outer conductor of which is deliberately made highly electrically resistive so that it can serve as a heating element. As in the cases of other axisymmetric coaxial cables, the equal-magnitude electric currents flowing in opposite directions along the inner and outer conductors give rise to zero net magnetic field outside the outer conductor. Hence, a coaxial electric heater can be placed near an atomic-sample cell or other sensitive device.

Figure 2. This Coaxial Electric Heater, shown herein a side view, was bent into a circular arc for edge heating of a circular window. This heater was made from copper wire of 0.01-in. (0.254-mm) diameter. Its resistance is about 2 kΩ.
A coaxial electric heater can be fabricated from an insulated copper wire, the copper core of which serves as the inner conductor. For example, in one approach, the insulated wire is dipped in a colloidal graphite emulsion, then the emulsion-coated wire is dried to form a thin, uniform, highly electrically resistive film that serves as the outer conductor. Then the film is coated with a protective layer of high-temperature epoxy except at the end to be electrically connected to the power supply. Next, the insulation is stripped from the wire at that end. Finally, electrical leads from the heater power supply are attached to the exposed portions of the wire and the resistive film.

The resistance of the graphite film can be tailored via its thickness. Alternatively, the film can be made from an electrically conductive paint, other than a colloidal graphite emulsion, chosen to impart the desired resistance. Yet another alternative is to tailor the resistance of a graphite film by exploiting the fact that its resistance can be changed permanently within about 10 percent by heating it to a temperature above 300 °C. Figure 2 depicts a coaxial heater, with electrical leads attached, that has been bent into an almost full circle for edge heating of a circular window. (In the specific application, there is a requirement for a heated cell window, through which an optical beam enters the cell.)

This work was done by Dmitry Strekalov, Andrey Matsko, Anatoliy Savchenkov, and Lute Maleki of Caltech for NASA's Jet Propulsion Laboratory.

NPO-43569



This Brief includes a Technical Support Package (TSP).