An effort is underway to develop improved biosensors of a type based on ion channels in biomimetic membranes. These sensors are microfabricated from silicon and other materials compatible with silicon. As described below, these sensors offer a number of advantages over prior sensors of this type.

To place these advantages in context, it is first necessary to present some background information on prior sensors of this type:

  • Ion channels of the type in question are very sensitive to a wide variety of ligands, to which they respond by generating specific electrical signals. Consequently, such ion channels would be ideal bases for biosensors if they could be constructed in arrays and made stable and electrically addressable.
  • Unfortunately, such ion channels function only within biological or biomimetic membranes, the life spans of which, heretofore, have usually been no more than several hours. In addition, their small signals (currents of 1 to 10 pA) make it necessary to mount them on platforms along with, and coupled to, lownoise, highly-amplifying electronic circuits. These characteristics have posed difficulties that have limited the commercial development of ion-channel biosensors.
  • Heretofore, the standard design of one commercial product line of ionchannel recording devices has provided for lipid-bilayer recording chambers. Such chambers are suitable for short-term (1 to 2 hours) recording only and are not constructed in arrays. Another commercial product line of ion-channel biosensors offers the stability needed for longer-term recording, but operation is limited to DC capacitive measurements.
An Array of Four Wells containing sensor structures was fabricated in a silicon wafer of 4-in. (≈10-cm) diameter to demonstrate the ability to make an array of multiple sensors on a single chip. Each sensor is electrically addressed, separately from the other sensors, via its electrodes.

The figure presents top and bottom views of a prototype array of four of the developmental sensors fabricated in a silicon wafer on a glass plate. Each sensor in the array can be individually electrically addressed, without interference with its neighbors. Each sensor includes a well covered by a thin layer of silicon nitride, in which is made a pinhole for the formation of lipid bilayer membrane. In one stage of fabrication, the lower half of the well is filled with agarose, which is allowed to harden. Then the upper half of the well is filled with a liquid electrolyte (which thereafter remains liquid) and a lipid bilayer is painted over the pinhole. The liquid contains a protein that forms an ionchannel on top of the hardened agarose. The combination of enclosure in the well and support by the hardened agarose provides the stability needed to keep the membrane functional for times as long as days or even weeks.

An electrode above the well, another electrode below the well, and all the materials between the electrodes together constitute a capacitor. What is measured is the capacitive transient current in response to an applied voltage pulse. One notable feature of this sensor, in comparison with prior such sensors, is a relatively thick dielectric layer between the top of the well and the top electrode. This layer greatly reduces the capacitance of an aperture across which the ion channels are formed, thereby increasing the signalto-noise ratio. The use of a relatively large aperture with agarose support makes it possible to form many ion channels instead of only one, thereby further increasing the signal-to-noise ratio and effectively increasing the size of the available ionic reservoir. The relatively large reservoir makes it possible to measure AC rather than DC.

This work was done by Jay Nadeau, Victor White, Dennis Dougherty, and Joshua Maurer of Caltech for NASA’s Jet Propulsion Laboratory.

In accordance with Public Law 96-517, the contractor has elected to retain title to this invention. Inquiries concerning rights for its commercial use should be addressed to:

Innovative Technology Assets Management
JPL
Mail Stop 202-233
4800 Oak Grove Drive
Pasadena, CA 91109-8099
(818) 354-2240
E-mail: This email address is being protected from spambots. You need JavaScript enabled to view it.

Refer to NPO-30710



This Brief includes a Technical Support Package (TSP).
Document cover
Improved Ion-Channel Biosensors

(reference NPO30710) is currently available for download from the TSP library.

Don't have an account? Sign up here.